TY - JOUR
T1 - Torsional Failure of Carbon Fiber Composite Plates Versus Stainless Steel Plates for Comminuted Distal Fibula Fractures
AU - Wilson, William K.
AU - Morris, Randal P.
AU - Ward, Adam J.
AU - Carayannopoulos, Nikoletta L.
AU - Panchbhavi, Vinod K.
N1 - Publisher Copyright:
© 2016 The Author(s).
PY - 2016/5
Y1 - 2016/5
N2 - Background: Carbon fiber composite implants are gaining popularity in orthopedics, but with few independent studies of their failure characteristics under supra-physiologic loads. The objective of this cadaveric study was to compare torsional failure properties of bridge plating a comminuted distal fibula fracture with carbon fiber polyetheretherketone (PEEK) composite and stainless steel one-third tubular plates. Methods: Comminuted fractures were simulated in 12 matched pairs of fresh-frozen human fibulas with 2-mm osteotomies located 3 cm proximal to the tibiotalar joint. Each fibula pair was randomized for fixation and implanted with carbon fiber and stainless steel 5-hole one-third tubular plates. The constructs were loaded in external rotation at a rate of 1-degree/sec until failure with a materials testing system. Torsional stiffness and mode of failure, as well as displacement, torque, and energy absorption for the first instance of failure and peak failure, were determined. Statistical analysis was performed with paired t tests and chi-square. Results: There were no significant differences among the 12 pairs for torsional stiffness, first failure torque, peak failure displacement, peak failure torque, or peak failure energy. Stainless steel plates exhibited significantly higher displacement (P <.001) and energy absorption (P =.001) at the first indication of failure than the carbon fiber plates. Stainless steel plates permanently deformed significantly more often than the carbon fiber plates (P =.035). Carbon fiber plates exhibited no plastic deformation with delamination of the composite, and brittle catastrophic failure in 1 specimen. Conclusions: In a comminuted human fibula fracture fixation model, carbon fiber implants exhibited multiple pre-peak failures at significantly lower angles than the first failure for the stainless steel implants, with some delamination of composite layers and brittle catastrophic failure rather than plastic deformation. Clinical relevance: The torsional failure properties of carbon fiber composite one-third tubular plates determined in this independent study provide novel in vitro data for this alternative implant material.
AB - Background: Carbon fiber composite implants are gaining popularity in orthopedics, but with few independent studies of their failure characteristics under supra-physiologic loads. The objective of this cadaveric study was to compare torsional failure properties of bridge plating a comminuted distal fibula fracture with carbon fiber polyetheretherketone (PEEK) composite and stainless steel one-third tubular plates. Methods: Comminuted fractures were simulated in 12 matched pairs of fresh-frozen human fibulas with 2-mm osteotomies located 3 cm proximal to the tibiotalar joint. Each fibula pair was randomized for fixation and implanted with carbon fiber and stainless steel 5-hole one-third tubular plates. The constructs were loaded in external rotation at a rate of 1-degree/sec until failure with a materials testing system. Torsional stiffness and mode of failure, as well as displacement, torque, and energy absorption for the first instance of failure and peak failure, were determined. Statistical analysis was performed with paired t tests and chi-square. Results: There were no significant differences among the 12 pairs for torsional stiffness, first failure torque, peak failure displacement, peak failure torque, or peak failure energy. Stainless steel plates exhibited significantly higher displacement (P <.001) and energy absorption (P =.001) at the first indication of failure than the carbon fiber plates. Stainless steel plates permanently deformed significantly more often than the carbon fiber plates (P =.035). Carbon fiber plates exhibited no plastic deformation with delamination of the composite, and brittle catastrophic failure in 1 specimen. Conclusions: In a comminuted human fibula fracture fixation model, carbon fiber implants exhibited multiple pre-peak failures at significantly lower angles than the first failure for the stainless steel implants, with some delamination of composite layers and brittle catastrophic failure rather than plastic deformation. Clinical relevance: The torsional failure properties of carbon fiber composite one-third tubular plates determined in this independent study provide novel in vitro data for this alternative implant material.
KW - biomechanics
KW - carbon fiber
KW - fibula fracture
KW - one-third tubular plates
KW - stainless steel
UR - http://www.scopus.com/inward/record.url?scp=84966332526&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84966332526&partnerID=8YFLogxK
U2 - 10.1177/1071100715625291
DO - 10.1177/1071100715625291
M3 - Article
C2 - 26755535
AN - SCOPUS:84966332526
SN - 1071-1007
VL - 37
SP - 548
EP - 553
JO - Foot and Ankle International
JF - Foot and Ankle International
IS - 5
ER -