Abstract
Locking unicortical screw fixation of diaphyseal fractures may be clinically necessary due to internal constructs or prosthetic implants. The biomechanical effects of substituting one or more bicortical nonlocking screws to create a "hybrid" construct are unknown. Use of an initial nonlocking screw may be clinically desirable for creating initial plate-to-bone fixation prior to locking. A validated transverse fracture model was used with eighteen synthetic humerus models. Plate and screw fixation was performed in three treatment groups (n = 6): Unicortical locking screws only, fixation with two unicortical locking screws and one bicortical nonlocking screw on either side of the fracture and fixation with one unicortical locking screw and two bicortical locking screws on either side of the fracture. Displacement and gap strain measurements were made with a cyclic 4-point bending test. No significant differences were noted for construct stiffness or gap micromotion among the three groups on cyclic testing. Measurements at 1000, 5000, and 10,000 cycles showed stable fixation of all construct types with no significant changes in stiffness or micromotion. Substitution of one or more bicortical nonlocking screws in a unicortical locked construct does not affect construct stiffness or fracture gap strain. Data proven through controlled laboratory study.
Original language | English (US) |
---|---|
Article number | 1450071 |
Journal | Journal of Mechanics in Medicine and Biology |
Volume | 14 |
Issue number | 5 |
DOIs | |
State | Published - Oct 2014 |
Externally published | Yes |
Keywords
- Biomechanical
- fixed angle
- hybrid plating
- locking plate
- unicortical
ASJC Scopus subject areas
- Biomedical Engineering