TY - JOUR
T1 - Early experience of a transcatheter aortic valve program at a veterans affairs facility
AU - Omer, Shuab
AU - Kar, Biswajit
AU - Cornwell, Lorraine D.
AU - Blaustein, Alvin
AU - Levine, Glen N.
AU - Ali, Nadir
AU - Jneid, Hani
AU - Paniagua, David
AU - Atluri, Prasad V.
AU - Bechara, Carlos F.
AU - Kougias, Panos
AU - Ruma, Maryrose
AU - Preventza, Ourania
AU - Bozkurt, Biykem
AU - Carabello, Blase A.
AU - Bakaeen, Faisal G.
PY - 2013/12
Y1 - 2013/12
N2 - IMPORTANCE The US Food and Drug Administration recently approved the use of a transcatheter aortic valve in patients for whom traditional valve replacement surgery poses a high or prohibitive risk. Our hospital was one of the first Veterans Affairs facilities to launch a transcatheter aortic valve replacement (TAVR) program. OBJECTIVE To evaluate our early experience with transfemoral TAVR. DESIGN AND SETTING We retrospectively reviewed the records of all patients who underwent TAVR during the first year of our program at the Department of Cardiothoracic Surgery, Michael E. DeBakey Veterans Affairs Medical Center. PARTICIPANTS The mean (SD) age of the patients was 77 (9) years, and their mean (SD) Society of Thoracic Surgeons predicted risk of mortality score was 8.8 (10.7). INTERVENTIONS All patients underwent TAVR with the SAPIEN transcutaneous valve. MAIN OUTCOME MEASURES We evaluated operative mortality and major operative morbidity (stroke, myocardial infarction, renal failure necessitating dialysis, and requirement for mechanical circulatory support, as well as vascular complications and requirement for permanent pacemaker), in addition to length of hospital stay and discharge status. RESULTS Between December 21, 2011, and December 13, 2012, a total of 19 transfemoral TAVR procedures were performed at our center. Implantation was successful in all cases. There were no reports of operative (30-day) mortality, prosthetic valve endocarditis, renal failure necessitating dialysis, perioperativemyocardial infarction or stroke, or conversion to surgical aortic valve replacement. Seven patients (37%) had mild paravalvular leak, 3 patients (16%) had moderate paravalvular leak, 2 patients (11%) had groin wound complications, 2 patients (11%) required a permanent pacemaker, 1 patient (5%) had a vascular access complication requiring endovascular repair, and 1 patient (5%) required temporary circulatory support (with extracorporeal membrane oxygenation). The mean (SD) length of hospital stay after TAVR was 8.0 (5.9) days. All patients were discharged home. CONCLUSIONS AND RELEVANCE Transcatheter aortic valve replacement can be performed safely and with good outcomes at a Veterans Affairs facility with a committed multidisciplinary team and substantial experience in heart valve and endovascular therapies.
AB - IMPORTANCE The US Food and Drug Administration recently approved the use of a transcatheter aortic valve in patients for whom traditional valve replacement surgery poses a high or prohibitive risk. Our hospital was one of the first Veterans Affairs facilities to launch a transcatheter aortic valve replacement (TAVR) program. OBJECTIVE To evaluate our early experience with transfemoral TAVR. DESIGN AND SETTING We retrospectively reviewed the records of all patients who underwent TAVR during the first year of our program at the Department of Cardiothoracic Surgery, Michael E. DeBakey Veterans Affairs Medical Center. PARTICIPANTS The mean (SD) age of the patients was 77 (9) years, and their mean (SD) Society of Thoracic Surgeons predicted risk of mortality score was 8.8 (10.7). INTERVENTIONS All patients underwent TAVR with the SAPIEN transcutaneous valve. MAIN OUTCOME MEASURES We evaluated operative mortality and major operative morbidity (stroke, myocardial infarction, renal failure necessitating dialysis, and requirement for mechanical circulatory support, as well as vascular complications and requirement for permanent pacemaker), in addition to length of hospital stay and discharge status. RESULTS Between December 21, 2011, and December 13, 2012, a total of 19 transfemoral TAVR procedures were performed at our center. Implantation was successful in all cases. There were no reports of operative (30-day) mortality, prosthetic valve endocarditis, renal failure necessitating dialysis, perioperativemyocardial infarction or stroke, or conversion to surgical aortic valve replacement. Seven patients (37%) had mild paravalvular leak, 3 patients (16%) had moderate paravalvular leak, 2 patients (11%) had groin wound complications, 2 patients (11%) required a permanent pacemaker, 1 patient (5%) had a vascular access complication requiring endovascular repair, and 1 patient (5%) required temporary circulatory support (with extracorporeal membrane oxygenation). The mean (SD) length of hospital stay after TAVR was 8.0 (5.9) days. All patients were discharged home. CONCLUSIONS AND RELEVANCE Transcatheter aortic valve replacement can be performed safely and with good outcomes at a Veterans Affairs facility with a committed multidisciplinary team and substantial experience in heart valve and endovascular therapies.
UR - http://www.scopus.com/inward/record.url?scp=84890714492&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84890714492&partnerID=8YFLogxK
U2 - 10.1001/jamasurg.2013.3743
DO - 10.1001/jamasurg.2013.3743
M3 - Article
AN - SCOPUS:84890714492
SN - 2168-6254
VL - 148
SP - 1087
EP - 1093
JO - JAMA Surgery
JF - JAMA Surgery
IS - 12
ER -