Abstract
Adherence of a biological graft to the wound surface is the most important factor influencing the ultimate success of graft viability. A machine has been developed to test the adherence of biological graft materials to a substrate such as a wound surface. The peeling mode, which yields reproducible quantitative measurements of adherence, is a standard method for testing adhesives. The device is designed to continuously measure the force required to peel the graft from the substrate at a constant rate. This force is a function of the energy of adhesion per unit area of adhered surface. This device has been used to measure the peeling force of (2×2 cm) skin grafts which are applied to full-thickness wounds on mice. Results of tests on adherence of autografts on mice show that the peeling force increases significantly with time over the first 9 days of healing. Thus, this device is useful in quantitative comparison of various skin grafting techniques and artificial grafts.
Original language | English (US) |
---|---|
Pages (from-to) | 51-55 |
Number of pages | 5 |
Journal | Annals of Biomedical Engineering |
Volume | 21 |
Issue number | 1 |
DOIs | |
State | Published - Jan 1993 |
Externally published | Yes |
Keywords
- Adherence
- Skin grafts
- Tensiometer
ASJC Scopus subject areas
- Biomedical Engineering