Abstract
BACKGROUND: Although accurate quantification of oxygen consumption (V˙ O2) and carbon dioxide elimination (V˙CO2) provides important insights into a patient’s nutritional and hemodynamic status, few devices exist to accurately measure these parameters in children. Therefore, we assessed the accuracy and agreement of 2 devices currently on the market using a pediatric in vitro model of gas exchange. METHODS: We utilized a Huszczuk simulation model, which simulates oxygen consumption and carbon dioxide production using gas dilution, to examine the accuracy of two FDA-cleared respiratory modules (E-COVX and E-sCAiOVX-00). V˙ O2 and V˙CO2 were set at 20, 40, 60, and 100 mL/min, ranges typical for infant and pediatric patients. Bland-Altman analysis was used to calculate the bias and limits of agreement of each device relative to simulated values for V˙ O2 and V˙CO2. RESULTS: The E-COVX mean percentage bias (limits of agreement) was =26.3% (=36.1 to =16.6%) and =39.3% (=47.5 to =31.1%) for V˙ O2 and V˙CO2, respectively. The mean bias (limits of agreement) for the E-aCAiOVX-00 was =0.5% (=13.3 to 12.3%) and =6.0% (=13.8 to 1.7%) for V˙ O2 and V˙CO2, respectively. CONCLUSIONS: The E-COVX demonstrated bias and limits of agreement that were not clinically acceptable; therefore, application of this module to pediatric patients would not be recommended. The new module, E-sCAiOVX, demonstrated acceptable bias and limits of agreement for the V˙ O2 and V˙CO2 in the range 40–100 mL/min (which corresponds to patients in the range of ∼5–16 kg).
Original language | English (US) |
---|---|
Pages (from-to) | 475-480 |
Number of pages | 6 |
Journal | Respiratory care |
Volume | 62 |
Issue number | 4 |
DOIs | |
State | Published - Apr 4 2017 |
Externally published | Yes |
Keywords
- Accu- racy
- Carbon dioxide elimination
- Indirect calorimetry
- Oxygen consumption
- Words: gas exchange
ASJC Scopus subject areas
- Pulmonary and Respiratory Medicine
- Critical Care and Intensive Care Medicine